The CellTram 4r Oil is a manual microinjector suitable for a broad range of life science applications. It is a hydraulic, oil-filled microinjector that provides more direct sample control and can generate higher pressures than an air system. It is optimized for leakage-free work and easy oil filling. The model can be used with all common micromanipulation systems yet it develops its full potential with Calibre Scientific manipulators. The CellTram 4r Oil injector is for research use only.
Applications
CSI of mouse and large mammals (e.g., equine, porcine...)
Holding of sensitive oocytes, embryos or small organisms
Piezo-assisted micromanipulation in combination with the Calibre Scientific PiezoXpert®
Transfer of cells or organelles (e.g., ES cell or iPS cell transfer into blastocysts or morulae)
Enucleation/nuclear transfer
(Xeno)transplantations
Removal of cells (e.g., for polar body biopsy, blastomere biopsy or trophectoderm biopsy)
Manual microinjection and dispensing of aqueous solutions
Single cell picking
Features
Improved oil-filling system for minimal interruption of the workflow
Coarse and fine drive for precise and intuitive sample control
Improved rotary knob design for better handling and easier cleaning
Two rotary knobs of different transmission ratio (1:1 and 1:10) for the highest standards in terms of resolution, sensitivity and response behavior
Capillary holder 4 with markings for easier positioning
New grip head 4 system for simple capillary change and higher user safety
Position indicators and marking for in/out to reduce risk of user errors
Stable, nonslip base plate
Highly reliable performance with minimal maintenance
QuickValve system for fast and simple filling
Compatible with most commonly used micromanipulators
Technical Data
Specifications
CellTram 4r Oil
Injection tube
Fluorinated ethylene propylene (FEP), 1.3 m length, inner diameter 0.5 mm, outer diameter 2 mm
Pressure generation
Cylinder/piston system, oil-filled
Volume change per revolution
1 µL/10 µL (fine / coarse rotary knob)
Cylinder volume
1000 µL
Min. adjustable volume
< 1.5 nL
Max. pressure
20000 hPa